| Maintenance
70
|
CO2 Incubator Basic
VWR
Chapter 10
Replacing the door seal
Figure 23 – Door seal replacement
The door seal (magnetic seal) of the outer door is located in the retaining slot. No tools are required
to replace the seal.
1. Pull magnetic seal [3] out of the guide slot [1].
2. Position new seal at a corner [2] and press seal retaining rail [4] into slot.
3. Make sure that the retaining rail taper is positioned correctly in the slot [1] and that the seal is
flush with the door frame.