| Installation of the device
14
|
CO2 Incubator Basic
VWR
Chapter 3
Therefore, the incubator must only be installed in rooms with sufficient ventilation.
Do not install the device in room recesses without ventilation.
The room ventilation should be a technical ventilation that is in accordance with the
requirements of DGUV Information 213-850 (Guidelines for laboratories) or some other
suitable ventilation system with appropriate capacity.
Space requirements
Figure 1 – Device dimensions
When installing the device, make sure that the installation and supply connections are freely
accessible.
The control box at the rear panel of the device may serve as a spacer to adjacent objects. The side
distances given are minimal distances.
To protect the CO
2
incubator against contamination, use a floor stand even if the device is installed
near the floor. The height of the floor stand should not be less than 200 mm.
Several floor stands and carriers are available as options from VWR (Part no. see Chapter 11
“Parts and accessories”).