| Start-up
32
|
CO2 Incubator Basic
VWR
Chapter 5
Levelling the device:
1. Position a bubble level on the center shelf.
2. Rotate the adjustable device stands using the supplied 24 mm wrench until the shelf is
positioned horizontally in all directions. Perform the adjustment of the device stands from left to
right and from rear to front.
Gas connection
NOTICE!
Gas quality:
CO
2
must be of the following qualities:
Purity 99.5% min
or
Medical grade
CAUTION
Overpressure:
The operating pressure of the gas applied to the device must not exceed 1 bar. If the gas is
supplied at a higher pressure, the valves integral to the device may not close correctly and the gas
supply control may be impaired!
Set the gas supply to a range between 0.8 bar min. and 1.0 bar max. and make sure that this
pressure setting cannot be changed!
NOTE!
Pressure compensation opening:
To ensure permanent pressure compensation, the pressure compensation opening must not be
connected to an exhaust air system. The pipe of the pressure compensation opening must not be
extended or redirected!
Access port:
If the access port is not used, it must be capped during operation!
Summary of Contents for CO2 Incubator Basic
Page 1: ...VWR CO2 Incubator Basic 50150229 Revision 0 December 2015 ...
Page 54: ... Shut down 52 CO2 Incubator Basic VWR Chapter 8 ...
Page 62: ... Cleaning and disinfection 60 CO2 Incubator Basic VWR Chapter 9 ...
Page 74: ... Parts and accessories 72 CO2 Incubator Basic VWR Chapter 11 ...
Page 78: ... Technical data 76 CO2 Incubator Basic VWR Chapter 12 ...
Page 83: ...VWR CO2 Incubator Basic 81 Certificate of decontamination ...
Page 84: ... Certificate of decontamination 82 CO2 Incubator Basic VWR Chapter 15 ...
Page 85: ......
Page 86: ......