Cleaning and disinfection
Decontamination procedures
The operator must prepare hygiene regulations for the decontamination of the device in
accordance with the application of the device.
Wipe/spray disinfection:
is used as the standardized manual disinfection procedure for the
device and for all accessories.
Wipe/Spray disinfection
The wipe/spray disinfection is carried out in three stages:
Predisinfection
Cleaning
Final disinfection
Recommended cleaning and disinfection agents:
NOTICE!
Incompatible cleaning agents:
Some device components are made of plastic. Solvents may dissolve plastics. Powerful acids or
bases may cause embrittlement of the plastics.
For cleaning the plastic components and surfaces, do not use hydrocarbon-containing solvents,
detergents with an alcohol content of more than 10% or powerful acids and bases!
Chloride-containing disinfectants:
Chloride-containing disinfectants may corrode stainless steel.
Use only disinfectants that do not affect stainless steel!
Do not use sodiumhypochlorite solutions such as Purex and Clorox. These can cause
corrosion and pitting of stainless steel.
Do not use steel wood pads such as Brillo. These pads deposit carbon particles in the work
space.
Summary of Contents for CO2 Incubator Basic
Page 1: ...VWR CO2 Incubator Basic 50150229 Revision 0 December 2015 ...
Page 54: ... Shut down 52 CO2 Incubator Basic VWR Chapter 8 ...
Page 62: ... Cleaning and disinfection 60 CO2 Incubator Basic VWR Chapter 9 ...
Page 74: ... Parts and accessories 72 CO2 Incubator Basic VWR Chapter 11 ...
Page 78: ... Technical data 76 CO2 Incubator Basic VWR Chapter 12 ...
Page 83: ...VWR CO2 Incubator Basic 81 Certificate of decontamination ...
Page 84: ... Certificate of decontamination 82 CO2 Incubator Basic VWR Chapter 15 ...
Page 85: ......
Page 86: ......