| Technical data
74
|
CO2 Incubator Basic
VWR
Chapter 12
HUMIDITY
Water quality
demineralized / distilled or
autoclaved or completely
deionized / distilled or
autoclaved.
Liquid quantity:
Incubation operation
90°C decontamination operation
l
ml
max. 3.0 / min 1.2
300
Constant humidity at 37 °C
% rH
approx. 93
Humidity recovery time, at 93% rH, door open 30 s
(to 95% of initial value)
min
min
approx. 20 (stainless steel)
CO
2
GAS SUPLLY SYSTEM
Gas purity
%
min. 99.5 or medical quality
Prepressure
bar
min. 0.8 - max. 1
Measuring and control range
% vol.
0... 20
Control deviation, temporal
% vol.
± 0.1
Recovery time, at 5%, door open 30 seconds
(to 98% of initial value)
min
< 8
CO
2
MEASURING CELL
Accuracy (absolute)
% CO
2
± 0.3
ELECTRICAL SYSTEM
Rated voltage
V
1/N/PE 120 V, AC
Rated frequency
Hz
50/60
Interference suppression (DIN VDE 0875)
Interference level N
Type of protection (DIN 40 050)
IP 20
Protection class
l
Overvoltage category (IEC 1010, EN 61010)
ll
Pollution severity (IEC 1010, EN 61010)
2
Rated current
A
5.3 (120 V, AC)
On-site fusing:
Fuse
Circuit breaker
T 16 A
G 16
Rated input
kW
0.64 (120 VAC)
EMC class
B
OTHERS
Sound pressure level (DIN 45 635, Part 1)
dB(A)
< 50
Relative humidity of environment
% rH
max. 80
Location elevation
m NN
max. 2000
TECHNICAL SPECIFICATIONS
DESCRIPTION
UNITS:
VALUE:
Summary of Contents for CO2 Incubator Basic
Page 1: ...VWR CO2 Incubator Basic 50150229 Revision 0 December 2015 ...
Page 54: ... Shut down 52 CO2 Incubator Basic VWR Chapter 8 ...
Page 62: ... Cleaning and disinfection 60 CO2 Incubator Basic VWR Chapter 9 ...
Page 74: ... Parts and accessories 72 CO2 Incubator Basic VWR Chapter 11 ...
Page 78: ... Technical data 76 CO2 Incubator Basic VWR Chapter 12 ...
Page 83: ...VWR CO2 Incubator Basic 81 Certificate of decontamination ...
Page 84: ... Certificate of decontamination 82 CO2 Incubator Basic VWR Chapter 15 ...
Page 85: ......
Page 86: ......