| Important safety information
10
|
CO2 Incubator Basic
VWR
Chapter 1
Standards and directives
The device is in accordance with the following standards and guidelines:CAN/
CSA – C22.2 No. 61010-4-04
UL Std. No. 61010 (2nd Edition)
Safety specifications:
Indoor Use Only
Altitude – Up to 2.000 meters
Temperature – 5 °C to 43 °C
Humidity – Maximum RH 80% for temperatures up to 31 °C, decreasing linearly to 50% RH at
40 °C
Mains Supply Fluctuations – Mains supply voltage fluctuations not to exceed ±10% of the
nominal voltage
Installation Category II
1
Pollution Degree 2
2
Class of Equipment I
1
Installation category (overvoltage category) defines the level of transient overvoltage which the instrument is designed
to withstand safely. It depends on the nature of the electricity supply and is overvoltage protection means. For example,
in Cat II which is the category used for instruments in installations supplied from a supply comparable to public mains
such a hospital and research laboratories and most industrial laboratories, the expected transient over voltage is 2500 V
for a 230 V supply and 1500 V for a 120 V supply.
2
Pollution degree describes the amount of conductive pollution present in the operating environment. Pollution degree 2
assumes that normally only non-conductive pollution such a dust occurs with the exception of occasional conductivity
caused by condensation.
Summary of Contents for CO2 Incubator Basic
Page 1: ...VWR CO2 Incubator Basic 50150229 Revision 0 December 2015 ...
Page 54: ... Shut down 52 CO2 Incubator Basic VWR Chapter 8 ...
Page 62: ... Cleaning and disinfection 60 CO2 Incubator Basic VWR Chapter 9 ...
Page 74: ... Parts and accessories 72 CO2 Incubator Basic VWR Chapter 11 ...
Page 78: ... Technical data 76 CO2 Incubator Basic VWR Chapter 12 ...
Page 83: ...VWR CO2 Incubator Basic 81 Certificate of decontamination ...
Page 84: ... Certificate of decontamination 82 CO2 Incubator Basic VWR Chapter 15 ...
Page 85: ......
Page 86: ......