41
Beat2Phone ECG Instruction Manual
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic envi-
ronment – guidance
Conducted RF
IEC 61000-4-6
150 kHz to 80 MHz
3 V
Portable and mobile RF
communications equip-
ment should not be used
any closer to any part of
the Beat2Phone ECG, in-
cluding cables, than the
recommended separation
distance calculated from
the equation applicable to
the frequency of the
transmitter.
Recommended sepa-
ration distance:
𝑑 = √𝑃 ∗ 1.17 𝑚
𝑑 = √𝑃 ∗ 1.17 𝑚
80 – 800
MHz
𝑑 = √𝑃 ∗ 2.33 𝑚
800
MHz – 2,7 GHz
where
P
is the maximum
output power rating of
the transmitter in watts
(W) according to the
transmitter manufacturer
and
d
is the recom-
mended separation dis-
tance in meters (m).
The strengths of fields
produced by fixed RF
transmitters, as deter-
mined by an electromag-
netic site survey,
a
should
be less than the compli-
ance level in each fre-
quency range.
b
Interfer-
ence may occur in the vi-
cinity of equipment
marked with the follow-
ing symbol:
ISM and amateur
radio bands between
0.15 MHz and 80 MHz
6 V
Radiated RF
IEC 61000-4-3
80 MHz to 2.7 GHz
10 V/m
Proximity fields from
wireless RF communica-
tions equipment
IEC 61000-4-3
Proximity fields @ ISM
bands according to:
Table 9
Table 9 of 60601-1-
2:2015
ESD EN 61000-4-2
Contact discharge (di-
rect/indirect)
8 kV
Air discharge (direct)
2, 4, 8, 15 kV
Power frequency mag-
netic fields EN 61000-4-
8
Power-frequency mag-
netic field, 30 A/m
Continuous field (50/60
Hz)
Power supply networks
voltage dips and inter-
ruptions EN 61000-4-11
Voltage reduction 100
%, 0.5, 1, 250 cycles
10, 20, 5000 ms
Voltage reduction 70%,
25 cycles
500 ms
Surge EN 61000-4-5
Line to line
+/- 0.5, 1.0 kV
Electrical fast transient
burst
EN 61000-4-4
Individual supply lines,
100 kHz rate
2 kV
All combinations of sup-
ply lines and earth, 100
kHz rate
NOTE
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorp-
tion
and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoreti-
cally with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromag-
netic site survey should be considered. If the measured field strength in the Beat2Phone ECG use location
exceeds the applicable RF compliance level above, the Beat2Phone ECG should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting
or relocating the Beat2Phone ECG
.
In the the frequency range 150 kHz – 80 MHz, field strengths should be
less than 3 V/m.
Recommended separation distances between portable and mobile RF communications equip-
ment and the Beat2Phone ECG