12
Beat2Phone ECG Instruction Manual
6.1
Attaching the ECG skin contact electrodes
Proper electrode placement and low impedance contact to skin are important
prerequisites for high quality ECG measurement. The electrode strap for the sin-
gle-lead ECG should be placed across the chest as depicted in
Figure 6-1Figure
6-1
. Adjust the electrode strap to achieve the best possible signal.
Figure 6-1: The Beat2Phone sensor properly attached to the chest with the electrode
strap.
Gently clean the skin under the electrode strap with a wet wipe before placing
the electrode strap on the skin. No electrode gel or paste is needed for skin con-
tact.
Moisten the electrode strap contacts with a damp cloth (moistened with water)
before using the strap. This improves the electrical contact between the skin and
electrodes.
WARNING:
Do not use the device if skin under the electrode strap is damaged,
irritated, sensitive or indicates an allergic reaction.
The electrode strap can cause skin irritation. It is important to clean the strap and
care for the skin between using the electrodes.
NOTE!
The electrode strap is only intended for personal use during a single 1-2
week monitoring period.
Sensor is sensitive to ESD, so the electrode strap shall be placed firmly on the
chest before turning on and connecting the sensor to the snap contacts. Before
removing the electrode strap from the chest, the sensor shall first be detached