Dosing system eco-CONTROL EC200 DUO
Operating and maintenance instructions
Version 1.4
Copyright ©
5.19
Bleeding the 2K-dispenser after filling / refilling it and after cleaning
The dispenser must be bled when it is used for the first time and each time after refilling
or cleaning. Do this in accordance with the notes given in the operating instructions of the
2K-dispenser.
It is best to use the
start / stop program
with a medium flow rate to control the dispenser.
Caution:
Do not switch on
the Control system until medium has been delivered to it.
Otherwise there is a risk of
damage to the equipment.
Even a
short
test run can
cause
irreparable damage to the stator.
5.20
Error messages
5.21
Clearing error messages
If an error message is present, a flashing symbol appears in the status bar and and
acoustic signal sounds.
(if not switched off, see 5.10.14, System and error messages, page 26)
• Press the
Info
key, the acoustic signal is switched off, and the corresponding error
message appears.
- Overcurrent shutdown! Check the pump and motor
- Check the minimum fill level
- Check the compressed air
- Fill level A (or B) minimum
• Correct the error and clear the error message with
OK
.
Caution:
If the error message
Overcurrent switching
appears, the 2K-dispenser must
be cleaned before any further use.
5.22
Error message dosing pressure monitor error
If an error is present in one of the two dosing pressure monitor sensors, an error
message appears in the display. The error-code number provided enables us to
describe the error exactly. Please tell us the number or replace the relevant sensor.
Summary of Contents for PreeFlow eco-CONTROL EC200 DUO