If the configuration includes more than one acquisition unit, verify that a unique IP address is assigned to
each. The Unit address designating the equipment must also be unique. (See section
).
5.4 Analog input/output modules
Verify the LED indicators at the rear of the analog input/output modules upon startup:
•
The LED indicators at the rear of the modules should light up Orange for 2 seconds upon startup, then
turn either Green or Red depending on the status of the measuring chain connected to the module.
•
In some instances, it is normal for the LED to briefly turn Green before switching to Red.
•
If the LED indicator at the rear of a module remains Orange, a firmware problem has been encountered
and new firmware must be reloaded into the acquisition unit. See section
3.5 Firmware update to version
.
Verify, on the front panel display, that the status of each measuring chain confirms the color of the LED
indicator color at the back of the corresponding analog input/output module. If a bar graph and a numerical
value are displayed on the front panel display, then the parameter is correctly read.
Otherwise, verify the message associated with each channel:
•
NotOK
: an out of range problem has occurred. The LED indicator at the rear of the analog input/output
module is Red. Verify the measuring chain installation.
•
NotOK(T)
: the configuration does not match the module type;
•
NotOK(F)
: the firmware version is incorrect. See section
3.5 Firmware update to version 7.4 with a
•
NotOK(A/D)
: the A/D card is not working. Contact VibroSystM.
•
NotOK(V)
: the configuration version does not match the software version;
•
NotOK (*)
: The module is not responding;
•
NotOK (C)
: there is a problem with the configuration;
•
Poles not OK
: rotor deformation cannot be monitored because the detected number of poles is
incorrect
•
Speed not OK
: rotor deformation cannot be monitored because the unit is not operating within the
required speed range (99% to 101% of nominal speed for hydroelectric alternators, or 1% to 200% of
nominal speed for mills);
•
Stabilizing
: the sensor is connected but waits for the signal to stabilize before displaying a value;
•
Waiting
: the sensor is connected and the signal has been stabilized but the server has not completed its
trending calculation cycle.
44
ZPU-5000 Acquisition and Processing Unit with ZOOM Version 7.4 - User’s Manual