1.3.2 ZPU-5000 unit - Rear view
Terminal block connectors with removable female screw terminal connectors allow permanent wiring of cable
harnesses with the advantage of quick connect/disconnect.
Format
Location
Recommended Wire Size
3-pos.
ZPU-5000 Power Input
1,5 mm
2
[16 AWG]
(300 V rating)
1-pos.
ZPU-5000 Protective Conductor Terminal
1,5 mm
2
[16 AWG]
, up to 6 mm
2
[10 AWG]
3-pos.
Analog I/O Modules
0,5 - 0,34 mm
2
[20 - 22 AWG]
14-pos.
Analog I/O Modules
0,5 - 0,34 mm
2
[20 - 22 AWG]
6-pos.
Communication Module
0,5 - 0,34 mm
2
[20 - 22 AWG]
4-pos.
Control Module
0,5 - 0,34 mm
2
[20 - 22 AWG]
5-pos.
Control Module
0,5 - 0,34 mm
2
[20 - 22 AWG]
7-pos.
Control Module
0,5 - 0,34 mm
2
[20 - 22 AWG]
9-pos.
Control Module
0,5 - 0,34 mm
2
[20 - 22 AWG]
1.3.2.1 Analog I/O modules
Up to eight dual-channel analog I/O modules can be present to receive and process signals from various
sensors and conditioners. Each channel also provides four analog outputs and four open-collector outputs for
controlling remote alarm relays.
1.3.2.2 Control module
One control module is dedicated to the management of the control and synchronization signals for the
ZPU-5000 unit. Related inputs and outputs include: