Specific Methods Of Creating ADFs
4200-002
80
Revision 6
Cellcorder records data from the same battery. For
example, on a 256
–cell battery, one Cellcorder tests cells 1
to 128 and stores the data in Cellcorder locations 1 to 128.
The other Cellcorder tests cells 129 to 256 and stores in 1
to 128 as well. Capture the first Cellcorder with the starting
cell number set to one and capture the second into the
same file with the starting number set to 129.
To start capturing, click
.
Figure 158. Warning
If information will be
overwritten, a
warning window
appears.
Press
to
continue.
Figure 159. Transferring Data
NOTE:
The Communication Status shows; Transferring
configuration|cell data.
The CRT displays Sending data for Cell n.
When done, Communication Status shows Data transfer
completed.