39
PART 3 - Technical manual
Routing actuator cables (including cable ties)
The actuator cables are routed along the inside
of the bed frame to the control unit.
The VENDLET systems are delivered with eight
grey cable ties - two for each actuator. Route
the cable to the back of the actuator shaft and
use two grey cable ties to attach it once the
actuator leg has been adjusted to the right
height.
Then attach the actuator cable to the bed
frame using cable ties. It is important that the
cables are routed tightly to ensure that they
will rest firmly against the bed frame; use at
least four cable ties on each side of the bed.
Connecting the hand control
Connect the hand control to the control unit in
the plug by this symbol
on the control unit.
The hand control should always hang on the
outside of the bed such as on the outside of
the headboard.
The hand control should be connected before
the actuator and power cables are connected.
In case you need to replace a faulty hand
control, begin by unplugging the power cable,
replace the hand control and then reconnect
the power cable to the mains outlet.
Cables attached using two grey cable ties
The actuator cable is fixed using cable ties
The hand control is connected to the
control unit