51
PART 3 - Technical manual
Replacing the actuator
1. Disconnect the power cable from the
mains outlet.
2. Disconnect the cable from the control
unit.
3. Disconnect the actuator cable from the
control unit and release it from the bed,
allowing it to hang freely.
4. Remove the 2 Allen screws securing the
actuator to the actuator fitting and
remove the actuator.
5. Remove the actuator cable from the
defective actuator if you want to re-use
the cable. The actuator cable is attached
to the actuator with a locking device.
Release the locking device by pressing a
small straight blade screwdriver against
the side of the locking device on both
sides.
6. Remove the actuator cable from the
defective actuator and install it on the
new actuator.
7. Mount the new actuator on the actuator
fitting and attach it using the two Allen
screws.
8. Secure the actuator cable with two new cable ties and route it along the inside
of the bed frame to the control unit. Use cable ties to secure the actuator cable
along the bed frame in such a way that the cable will not get caught by the
moving parts of the bed. See section ” Routing actuator cables (including cable
ties)” for further details.
9. Reconnect the power cable to the mains outlet.
10. Reset the system by simultaneously pressing the two top buttons on the hand
control. You will hear a series of rhythmic beeps. Press and hold the buttons until
the beeping stops.
11. Check whether the actuator is working properly.
Remove the Allen screws and free the
actuator
Release the locking device