PART 3 - Technical manual
Mounting the bars
Use two 8 x 35 mm button head bolts.
The red bar must be placed to the right-hand
side of the patient, corresponding to the
actuator fitting with a green O ring on the
actuator plug.
The blue bar must be placed to the left-hand
side of the patient, corresponding to the
actuator fitting with a yellow O ring on the
actuator plug.
The bar end with the motor cable must be
facing the headboard. Mount the bar on the
actuator fittings and tighten the button head
bolts manually.
It should be possible to move the motor
bracket on the actuator fitting which is why
the nut should be tightened to the point only
when the bolt meshes with the actual lock
component.
Adjustment of the height of the actuator fittings
The bars must be mounted in such a way that
their tops are 22 – 23 cm above the surface of
the mattress. This is achieved by releasing the
adjustment screw on the back of the mounting
bracket and elevating the actuator fitting to
the desired height.
The fitting has 13 adjustment holes,
corresponding to a total adjustment range
of 15.5 cm.
Please note that all 4 actuator fittings must be
mounted at the same height.
Apply pressure to the adjustment
screw to ensure its full insertion.
Blue bar to the left-hand side of the
patient; red bar to the right-hand side of
the patient
Adjustment holes specified on the back of
the mounting bracket