PART 3 - Technical manual
TROUBLESHOOTING
Problem
Checklist
The VENDLET does not work
Has the power been turned on?
Is the green diode on the control unit under the bed illuminated?
Is the green diode on the hand control illuminated?
Are all cables properly mounted and intact?
Replace the hand control.
Reset the system by simultaneously pressing the two top buttons
on the hand control. You will hear 10 rhythmic beeps. Press and
hold the buttons until the beeping stops.
Contact a qualified technician or Healthcare Lifting Specialist.
One of the bars does not work.
Is the cable from the bar intact and correctly mounted?
Switch the plugs of the two bars. If the same bar still does not
work, then the motor of that bar is defective. If the other bar is
not working, then the hand control or the control unit is
defective. Replace the hand control. If this does not solve the
problem, replace the control unit.
Reset the system by simultaneously pressing the two top buttons
on the hand control. You will hear a series of rhythmic beeps.
Press and hold the buttons until the beeping stops.
Replace the bar motor.
Contact a qualified technician or Healthcare Lifting Specialist.
One or both bars will not work
when a load is applied
Does the sheet runs over the bar, cf. page 25 of the user manual?
If the sheet runs under the bar, the motor disconnects when used.
This is a safety feature.
Switch the plugs of the two bars. If the same bar still does not
work, then the motor of that bar is defective. If the other bar is
not working, then the hand control or the control unit is
defective. Replace the hand control. If this does not solve the
problem, replace the control unit.
Replace the bar motor.
Contact a qualified technician or Healthcare Lifting Specialist.