46
PART 3 - Technical manual
Electrical parts
Control unit no.:
Check the control unit including plugs and mounting fittings for signs of damage.
Check whether the timer deactivates after 90 seconds of idle time.
Transformer no.:
Hand control no.:
Check the transformer for signs of damage.
Check the hand control and mounting
fixtures for signs of damage.
Check all cables, plugs, and relief cables for
signs of damage.
Check the functioning of all buttons and
whether the green diode is illuminated
when the system is active.
Check whether the green diode is illuminated.
Check all plugs, cables, and relief cables.
When using an extension cable, check whether
the cable has the necessary quadrant
(1.0 mm2).
Check the stop function.
Bars
Red side
Blue side
Check the motor for signs of damage.
Check all cables, plugs, and relief cables for signs of
damage.
Check whether the bar may be operated manually
when the motor is disconnected from the control
unit.
Check whether the motor has full pulling power
when you press the
/
and reduced pulling
power when you press the
/
by stopping the
bar with your hand.
Actuators
Check the plastic covers for signs of damage.
Check all cables, plugs, and relief cables for damage.
Check whether the cables are properly secured.
Check whether the cables are properly secured to
the bed frame with cable ties.
Check the slide function – lubricate, if necessary.