7 Set up with the display and adjustment module
VEGAFLEX 82 • 4 … 20 mA/HART - two-wire
44222-EN-170220
The additional current output cannot be used as an output in the
sense of a safety-instrumented application (SIL).
The procedure corresponds to the previous settings of the standard
current output. See "
Setup - Current output
".
In the main menu point "
Display
", the individual submenu points
should be selected one after the other and provided with the correct
parameters to ensure optimum adjustment of the display options. The
procedure is described in the following.
The following submenu points are available:
The submenu points are described below.
This menu item enables the setting of the requested national lan-
guage.
In delivery status, the sensor is set to the ordered national language.
In this menu item, you define the indication of the measured value
on the display. You can display two different measured values. In this
menu item, you define measured value 1.
The default setting for the displayed value 1 is "
Filling height Level
".
In this menu item, you define the indication of the measured value
on the display. You can display two different measured values. In this
menu item, you define measured value 2.
The default setting for the displayed value 2 is the electronics tem-
perature.
In this menu item, you define the display format of the measured
value on the display. You can define different display formats for the
two measured values.
You can thus define the number of decimal positions the measured
value is displayed with.
Display
Display - Menu language
Display - Displayed value
1
Display - Displayed value
2
Display - Display format