8. OPERATING ERRORS AND TROULE SHOOTING 

 

Problem 

Cause 

► 

Solution 

1.

 

No timer display on handpiece. 

The unit is in sleep mode.   

► 

Press the on/off button to wake the unit. 

2.

 

Can not activate/wake the light. 

Battery is drained.   

► 

Please recharge battery, and try again. 

Handpiece is out of order. 

► 

If not sure what happen, contact the dealer for 
further inspection. 

3.

 

Error code “E2” shows up and buzzer 
beeps 5 short tones. 

Handpiece is overheat 

► 

Please wait for minutes to have handpiece cooled 
down.   

► 

Waiting time can be shortened if handpiece is cooled 
by compressor air. 

4.

 

Error code “E1” shows up and buzzer 
beeps 3 short tones. 

Low battery voltage 

► 

Battery is drained. Please place handpiece back to 
charging base for recharging. 

5.

 

Charging indicator on handpiece is not 
on when placing handpiece on the 
charging base. 

Check the power plug connects to power outlet socket 
firmly. 

► 

If not sure what happen, contact the dealer. 

6.

 

Battery working time is significantly 
shortened. 

► 

Battery is exhausted after period use (generally 1 
year). Please contact the dealer to replace with a 
new battery. 

► 

If not sure what happen, contact the dealer for 
further information. 

7.

 

Whenever press the on/off button or 
mode button on keypad, there is no 
response in timer display, nor light. 

The light program is crashing. 

► 

Use a clip to poke through the hole on the back of 
handpiece to reset the light program. (Please refer to 
Section 9, figure 9) 

 
If the above steps do not solve the problem, contact the dealer or point of purchase. 
 
Note:
 We and our authorized distributors will make available on request circuit diagram, component 

part lists and other information to assist user's appropriate technical personnel to repair the 
light cure units which are designated by us as repairable.

 

Summary of Contents for LED-P

Page 1: ...S 19 01 1125 UV101 V1 3 Vector LED Curing Light Vector LED P Instructions For Use VECTOR R D INC 6824 19TH ST 230 UNIVERSITY PLACE WA 98466 USA Website www vectorusa net User Manual...

Page 2: ...e Environment 6 Installation of the unit p 7 6 1 Initial Setup 6 2 Keypad And Modes 6 3 Timer Selection 6 4 Battery Charging 7 Operation Instruction p 10 8 Operating Errors Trouble shooting p 11 9 Ser...

Page 3: ...ON Do not look directly into the light emitted from this curing unit It may be harmful to the eyes Exposure must be restricted to the area of the oral cavity in which clinical treatment is intended Do...

Page 4: ...Check unit for adequate light output before each procedure Failure to verify output may allow inadequate curing 6 In order to avoid electric shock do not introduce any objects into the unit with the...

Page 5: ...t output at about 1 400 mW cm 3 Seconds to cure most resin based composite in the market It also features Digital timer display multiple mode selection boost fast curing and ramp up soft start Two but...

Page 6: ...25 minutes Dimensions Diameter 25 30 mm Length with without light guide 240 mm 180 mm Weight with light guide and battery Roughly 155 g 5 3 Charging Base and Handpiece Time to charge empty battery App...

Page 7: ...take about 6 hours Use only the power supply furnished with the light system Attempting to use another power supply source may cause injury or damage and will void the warranty 3 Mode button surround...

Page 8: ...all local requirements Always use unit in conjunction with a properly grounded outlet 6 2 Keypad and modes There are 2 buttons on the key pad for the mode timer selection and the on off button Fig 5...

Page 9: ...tached to the charging base or attach the power supply directly to the handpiece When this is done the POWER INDICATOR On Off button surrounding will illuminate Figure 7 The light may be used with the...

Page 10: ...the surface being treated Rotate the light guide into the optimal position for polymerization Place the light guide as close to the filling material as possible Avoid directly contacting the filling m...

Page 11: ...when placing handpiece on the charging base Check the power plug connects to power outlet socket firmly If not sure what happen contact the dealer 6 Battery working time is significantly shortened Ba...

Page 12: ...l accessories Part Description Contents LM200 Digital Light Radiometer 200 1x Light meter 200 810 Sterilizable fiber optic light guide 10mm diameter 1x 10mm diameter 809 Sterilizable fiber optic light...

Page 13: ...moved with alcohol a plastic spatula may help in removing the material Do not use any sharp or pointed tools to protect the surface of the light guide from scratching To disinfect place the cleaned li...

Page 14: ...re shield with a soft cloth and if required a mild cleaning agent Solvents or abrasive cleaners may not be used in any case as they damage the plastic components Make sure that charge connector remain...

Page 15: ...t by an authorized engineer dealer warranty will expire immediately 13 3 Warranty Exception The warranty stated herein is the sole warranty applicable to Vector R D products Vector R D expressly discl...

Reviews: