10. Cleaning / Disinfecting / Sterilizing

 

 
10.1 Cleaning and Disinfection of Fiber Optic Light Guide 

► 

Clean and disinfect the light guide before every use. The light guide is not sterile when delivered and 
must be sterilized before being used for the first time. 
 

► 

Please check the manufacturer’s information about the cleaning and disinfecting agents. 

► 

Make sure that the cleaning and disinfectant agents you have chosen do not contain any of the 
following materials: 

 

  Organic, mineral, and oxidizing acids (minimum acceptable pH value 5.5) 

 

  Bases (maximum acceptable pH value 8.5) 

 

  Oxidation agents (e. g., hydrogen peroxide) 

 

  Halogens (chlorine, iodine, bromide) 

 

  Aromatic/halogenized hydrocarbons 

► 

The light guide must not be exposed to temperatures higher than 134 °C (273 °F). 

 

► 

Use a soft brush or a soft cloth to manually remove gross contaminations. Adhering polymerized 
composite should be removed with alcohol, a plastic spatula may help in removing the material. Do 
not use any sharp or pointed tools to protect the surface of the light guide from scratching. 

 

► 

To disinfect, place the cleaned light guide for the specified application time into the solution, making 
sure that it is completely covered. Disinfectants containing ophthalaldehyde are recommended. 

► 

Remove the light guide from the solution and rinse thoroughly (at least 10 sec.) in water with low 
germ count. 

► 

Dry the light guide with a clean cloth. 

 
Sterilization 
Effective cleaning and disinfection are absolutely essential requirements for effective sterilization. Only 
steam sterilization is approved as a sterilization procedure: 

► 

Maximum sterilization temperature 134 °C (273 °F) 

► 

Sterilization time (exposure time at sterilization temperature) at least 20 min. at 121°C (250 °F) or 
at least 3 min. at 132 °C (270 °F) /134 °C (273 °F) 

 
Check 
Before using the light guide again, check it for damaged surfaces, discoloration, and contamination; do 
not use damaged light guides. If the light guide is still contaminated, repeat the cleaning and 
disinfection. 
 

Summary of Contents for LED-P

Page 1: ...S 19 01 1125 UV101 V1 3 Vector LED Curing Light Vector LED P Instructions For Use VECTOR R D INC 6824 19TH ST 230 UNIVERSITY PLACE WA 98466 USA Website www vectorusa net User Manual...

Page 2: ...e Environment 6 Installation of the unit p 7 6 1 Initial Setup 6 2 Keypad And Modes 6 3 Timer Selection 6 4 Battery Charging 7 Operation Instruction p 10 8 Operating Errors Trouble shooting p 11 9 Ser...

Page 3: ...ON Do not look directly into the light emitted from this curing unit It may be harmful to the eyes Exposure must be restricted to the area of the oral cavity in which clinical treatment is intended Do...

Page 4: ...Check unit for adequate light output before each procedure Failure to verify output may allow inadequate curing 6 In order to avoid electric shock do not introduce any objects into the unit with the...

Page 5: ...t output at about 1 400 mW cm 3 Seconds to cure most resin based composite in the market It also features Digital timer display multiple mode selection boost fast curing and ramp up soft start Two but...

Page 6: ...25 minutes Dimensions Diameter 25 30 mm Length with without light guide 240 mm 180 mm Weight with light guide and battery Roughly 155 g 5 3 Charging Base and Handpiece Time to charge empty battery App...

Page 7: ...take about 6 hours Use only the power supply furnished with the light system Attempting to use another power supply source may cause injury or damage and will void the warranty 3 Mode button surround...

Page 8: ...all local requirements Always use unit in conjunction with a properly grounded outlet 6 2 Keypad and modes There are 2 buttons on the key pad for the mode timer selection and the on off button Fig 5...

Page 9: ...tached to the charging base or attach the power supply directly to the handpiece When this is done the POWER INDICATOR On Off button surrounding will illuminate Figure 7 The light may be used with the...

Page 10: ...the surface being treated Rotate the light guide into the optimal position for polymerization Place the light guide as close to the filling material as possible Avoid directly contacting the filling m...

Page 11: ...when placing handpiece on the charging base Check the power plug connects to power outlet socket firmly If not sure what happen contact the dealer 6 Battery working time is significantly shortened Ba...

Page 12: ...l accessories Part Description Contents LM200 Digital Light Radiometer 200 1x Light meter 200 810 Sterilizable fiber optic light guide 10mm diameter 1x 10mm diameter 809 Sterilizable fiber optic light...

Page 13: ...moved with alcohol a plastic spatula may help in removing the material Do not use any sharp or pointed tools to protect the surface of the light guide from scratching To disinfect place the cleaned li...

Page 14: ...re shield with a soft cloth and if required a mild cleaning agent Solvents or abrasive cleaners may not be used in any case as they damage the plastic components Make sure that charge connector remain...

Page 15: ...t by an authorized engineer dealer warranty will expire immediately 13 3 Warranty Exception The warranty stated herein is the sole warranty applicable to Vector R D products Vector R D expressly discl...

Reviews: