6 Acquiring PANO Images
Aligning the Laser Beams
WARNING
Be careful not to project the laser beams directly into the
patient’s eyes as this could severely damage the patient’s
vision.
A.
Mid-sagittal plane laser beam
: Position the Mid-sagittal plane laser beam
at the center of the patient’s face to prevent magnifications of the left or right
side in the final image.
B.
Frankfurt plane laser beam
: Position the patient’s head so that the
Frankfurt plane is aligned with the Frankfurt plane laser beam.
* The Frankfurt plane is a plane which joins the lower eye lid to the
superior border of the external auditory meatus.
Frankfurt plane laser
beam UP button
Frankfurt plane laser
beam DOWN button
Frankfurt plane laser
beam
Mid-sagittal plane
laser beam