PaX-i3D Green (PHT-60CFO) User Manual
79
6 Acquiring PANO Images
F. The patient’s gender will be automatically selected according to the patient
information registered in EasyDent / EzDent-i. Nevertheless, ensure that this
information is correct.
G. Select X-ray intensity.
NOTE
Depending on the circumference of the patient’s head,
X-ray intensity may be classified as Hard, Normal, or Soft :
Soft ≤ Normal ≤ Hard
Age
Group
Average head
circumference
Range
(cm)
classification of head
circumference
Child
53
±
3 cm
> 53
±
3
Hard
53
±
3
Normal
53
±
3 <
Soft
Adult
56
±
3 cm
> 56
±
3
Hard
56
±
3
Normal
56
±
3 <
Soft
H. A default value for the tube voltage (kVp) and current (mA) will be displayed
based on the gender and X-ray intensity. If necessary, further adjustments can
be made using the arrows to the right of each number.
NOTE
Adjustable resolution:
-
Tube voltage: ± 1 kVp
-
Tube current: ± 1 mA
I. Click the
CONFIRM
button for these parameters to take effect.
NOTE
Please wait for a moment, while the rotating unit moves to
its initial scanning position.
[PI3DG_130U_44A_en]User Guide.indd 79
2016-05-24 오후 4:18:55
Summary of Contents for PaX-i3D Smart
Page 1: ...Version 1 5 0 PHT 60CFO User Manual E n g l i s h...
Page 2: ......
Page 3: ......
Page 4: ......
Page 10: ......
Page 14: ......
Page 42: ...This page is left intentionally blank...
Page 110: ......
Page 126: ......
Page 127: ...Troubleshooting 9...
Page 133: ...Disposing of the Unit 11...
Page 146: ......
Page 167: ...13 Samsung 1 ro 2 gil Hwaseong si Gyeonggi do Korea Postal Code 18449...