PaX-i3D Green (PHT-60CFO) User Manual
103
7 Acquiring CEPH Images
E. Ask the patient to stand upright facing
the sensor.
Make sure that the patient’s shoulders are
level and that his/her neck is relaxed.
F. Adjust the height of the unit to suit the
patient by pressing the column up/down
button or switch (optional).
CAUTION
After adjusting the column to the patient’s height, fit the ear
rods along the patient’s ear canals.
G. The ear rods should fit properly along
the patient’s ear canals. The patient’s
Frankfurt Plane should be parallel to the
floor.
H. Ask the patient to swallow any saliva in his/her mouth and remain still with
his/her mouth closed until X-ray exposure is complete.
I. Click the
READY
button after the patient has been positioned. No X-ray will
be emitted at this point.
J. Proceed to section
7.3 Initiating X-ray Exposure.
Frankfurt plane
[PI3DG_130U_44A_en]User Guide.indd 103
2016-05-24 오후 4:19:05