7 Acquiring CEPH Images
NOTE
The Multi FOV is supported for the equipment with the one
shot type sensor.
A. Click the
CEPH
button.
B. Select the scanning mode under
CEPH Examination. – Lateral, Full
Lateral, PA, SMV, Waters View, Carpus.
C. Select the FOV size for the specific imaging mode.
[With one shot type sensor]
FOV
Details
12 x 10 (inches)
30.48x25.40 (cm)
Full size
9 x 10 (inches)
22.86x25.40 (cm)
Region of the no interest of the rear part of the
head is eliminated to minimize the X-ray exposure
area.
8 x 8 (inches)
20.32x20.32 (cm)
Child
D.
EXPOSURE TIME
NOTE
Exposure time can be adjusted by resolution of 0.1 s in the
range of 0.5 s to 1.2 s.
For information regarding to the exposure time of each
CEPH imaging mode, refer to the section “4.3.2 CEPH”.
SC (Xmaru2301CF Sensor
Scan Type)
①
②