6 Acquiring PANO Images
6.3.1
PANO Standard and Bitewing mode
Positioning the patient
A. Insert the bite block into the Chinrest
receptacle on the equipment.
B. Place a hygienic cover over the bite
block.
WARNING
The hygiene cover for the bite block and chin support is
intended for single use only. Always replace the hygiene
cover for each new patient.
CAUTION
Disinfect the Chinrest, Chin support and bite block using
an alcohol-based cleaning solution and wipe away all
residue with dry cloth.
C. Open the temple supports by clicking
the temple support adjustment button.
D. Guide the patient —facing the Chinrest — to the equipment.
E. Adjust the height of the column using the column up/down button or switch
(optional) until the patient’s chin rests comfortably on the Chinrest.
Bite block
Hygiene cover
*To link to the video
m a n u a l : s c a n t h e
QR code with smart
phone or pad.
Lock /
Release knob
for bite block
[PI3DG_130U_44A_en]User Guide.indd 82
2016-05-24 오후 4:18:57