Appendix
EzRay Air Wall (VEX-S300W) User Manual
59
EN
G
LISH
A.2 Tables of Exposure Times (Default)
The following exposure timetables were established with a unit equipped with a cone
that corresponds to a focus-to-skin distance of 200 mm (8 inches), respectively.
65 kV, 3.0 mA
Receptor
Patient
Teeth
Angle of
inclination
SSD: 200 mm (8 inches)
kV
mA
s
Sensor
Adult
Incisor
Maxilla: +45
˚
Mandible: -25
˚
65
3.0
0.14
Canine
Maxilla: +45
˚
Mandible: -20
˚
65
3.0
0.16
Molar/
Premolar
Maxilla: +30
˚
Mandible: -5
˚
65
3.0
0.18
Bitewing
+5
˚
~ +8
˚
65
3.0
0.19
Child
Incisor
Maxilla: +45
˚
Mandible: -25
˚
65
3.0
0.11
Canine
Maxilla: +45
˚
Mandible: -20
˚
65
3.0
0.13
Molar/
Premolar
Maxilla: +30
˚
Mandible: -5
˚
65
3.0
0.15
Bitewing
+5
˚
~ +8
˚
65
3.0
0.16
Summary of Contents for EzRay Air Wall
Page 1: ...English User Manual Model VEX S300W Version 1 38 Full version...
Page 2: ......
Page 3: ......
Page 4: ......
Page 5: ......
Page 7: ...Notice vi EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Page 27: ...3 System Overview 18 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Page 47: ...4 Operation 38 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Page 51: ...6 User Maintenance 42 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Page 57: ...9 Product Specifications 48 EzRay Air Wall VEX S300W User Manual...
Page 58: ...9 Product Specifications EzRay Air Wall VEX S300W User Manual 49 ENGLISH...
Page 63: ...9 Product Specifications 54 EzRay Air Wall VEX S300W User Manual 4 Tube Dimensions mm...
Page 83: ......
Page 84: ......
Page 85: ......