Appendix
EzRay Air Wall (VEX-S300W) User Manual
61
EN
G
LISH
2. Leakage Dose
Scope
IEC 60601-2-65 203.12.4
Requirements
In the LOADING STATE, the AIR KERMA due to LEAKAGE RADIATION from X-RAY
SOURCE ASSEMBLIES, 1 m from the FOCAL SPOT, average over an area of 100
cm
2
of which no principal linear dimension exceeds 20 cm when operated at the
NOMINAL X-RAY TUBE VOLTAGE under the condition of LOADING corresponding to
the reference LOADING conditions, shall not exceed 0.25 mGy in one hour.
Leakage Dose
Permissive Range
65 kVp, 3.0 mA, 0.5 s (Max. Exposure Condition)
At Focal Spot to Distance 1 m
Duty Cycle 1 : 60
< 0.25 mGy/h
Test Equipment
Instrument
Manufacturer
Model
S/N
X / Gamma
Survey Meter
Radcal Co.
9015/10X5-180
91-1470/19069
Summary of Contents for EzRay Air Wall
Page 1: ...English User Manual Model VEX S300W Version 1 38 Full version...
Page 2: ......
Page 3: ......
Page 4: ......
Page 5: ......
Page 7: ...Notice vi EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Page 27: ...3 System Overview 18 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Page 47: ...4 Operation 38 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Page 51: ...6 User Maintenance 42 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Page 57: ...9 Product Specifications 48 EzRay Air Wall VEX S300W User Manual...
Page 58: ...9 Product Specifications EzRay Air Wall VEX S300W User Manual 49 ENGLISH...
Page 63: ...9 Product Specifications 54 EzRay Air Wall VEX S300W User Manual 4 Tube Dimensions mm...
Page 83: ......
Page 84: ......
Page 85: ......