Appendix
62
EzRay Air Wall (VEX-S300W) User Manual
Results
The following exposure timetables were established with a unit equipped with a cone
that corresponds to a focus-to-skin distance of 200 mm (8 inches), respectively. When
the leakage doses have been measured with each cover type (default, rectangular
2x3, and rectangular 4x3), all the results have been ND (Not Detected).
The raw data about the results are shown in the table below.
Result (Horizontal Plane)
Direction
Default type
Rectangular 2x3
Rectangular 4x3
[mGy/h]
[mGy/h]
[mGy/h]
0°
ND
ND
ND
20°
ND
ND
ND
45°
ND
ND
ND
65°
ND
ND
ND
90°
ND
ND
ND
110°
ND
ND
ND
135°
ND
ND
ND
155°
ND
ND
ND
0
0.2
0.4
0.6
0.8
1
1.2
0°
20°
45°
65°
90°
110°
135°
155°
180°
200°
225°
245°
270°
290°
315°
335°
Summary of Contents for EzRay Air Wall
Page 1: ...English User Manual Model VEX S300W Version 1 38 Full version...
Page 2: ......
Page 3: ......
Page 4: ......
Page 5: ......
Page 7: ...Notice vi EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Page 27: ...3 System Overview 18 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Page 47: ...4 Operation 38 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Page 51: ...6 User Maintenance 42 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Page 57: ...9 Product Specifications 48 EzRay Air Wall VEX S300W User Manual...
Page 58: ...9 Product Specifications EzRay Air Wall VEX S300W User Manual 49 ENGLISH...
Page 63: ...9 Product Specifications 54 EzRay Air Wall VEX S300W User Manual 4 Tube Dimensions mm...
Page 83: ......
Page 84: ......
Page 85: ......