Results
The following exposure timetables were established with a unit equipped with a cone
that corresponds to a focus-to-skin distance of 200 mm (8 inches), respectively.
Test Equipment
Instrument
Manufacturer
Model
S/N
X / Gamma
Survey Meter
Radcal Co.
9015/10X5-180
91-1470/19069
Method
PMMA Phantom aligned to 280 mm away from Focal Spot
(with Position Indicating Device (80 mm))
65 kVp, 3.0 mA, 0.5 s (Max. Exposure Condition)
Measure point: 2,000 mm from PMMA Phantom
Result (Horizontal Plane)
Direction [°]
[μ
Gy/h]
0°
11.41
20°
3.90
45°
3.97
65°
2.24