9. Product Specifications
52
EzRay Air Wall (VEX-S300W) User Manual
9.2
Technical Specifications
X-ray Generator
Item
Description
High Voltage
Generator
(Assembly)
Model
DG- S0102V1
Rated output power
Max. 0.2 kW
Duty Cycle
1:60 or more
(Exposure time: Interval time)
Cooling Protection
Thermistor ≥ 65
°C
Inherent Filtration
1.8 mm Al / 65 kV
Total Filtration
Min. 1.5 mm Al
Type
Inverter Type
Tube Voltage
55-65 kV
Tube Current
1.0-3.0 mA
X-ray Tube
Manufacturer
VATECH Co., Ltd.
Model
V1-650304 (Stationary Anode type)
Version
2.0
Focal spot size
0.4 mm (IEC 60336)
Anode heat contents
Max. 2.7 kJ
Maximum Anode Heat
Dissipation
200 W
Target Material
Tungsten
Target Angle
12.5°
Inherent Filtration
Min. 1.5 mm Al
X-ray Coverage
70 mm at SID 200 mm
Tube Voltage
Max. 65 kV
Tube Current
Max. 3.0 mA
Summary of Contents for EzRay Air Wall
Page 1: ...English User Manual Model VEX S300W Version 1 38 Full version...
Page 2: ......
Page 3: ......
Page 4: ......
Page 5: ......
Page 7: ...Notice vi EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Page 27: ...3 System Overview 18 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Page 47: ...4 Operation 38 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Page 51: ...6 User Maintenance 42 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Page 57: ...9 Product Specifications 48 EzRay Air Wall VEX S300W User Manual...
Page 58: ...9 Product Specifications EzRay Air Wall VEX S300W User Manual 49 ENGLISH...
Page 63: ...9 Product Specifications 54 EzRay Air Wall VEX S300W User Manual 4 Tube Dimensions mm...
Page 83: ......
Page 84: ......
Page 85: ......