Installation, Operating & Maintenance Instructions
Series 615 DN 40 (I.D. 1.5”), RS485
VAT Vakuumventile AG, CH-9469 Haag, Switzerland
Tel +41 81 771 61 61 Fax +41 81 771 48 30 [email protected] www.vatvalve.com
289111EB
2013-02-05
85/96
Description
Required tool
35. Tighten plate screws to block.
36. Loosen the plat screws a quarter turn
counter clockwise so that plate is still
movable at shaft.
Allen Wrench
3mm
37. Press the plate (
A
) into valve seat.
38. lever Press the (
B
) by hand to axis (
C
).
Make sure that lever (
B
) is engaged at
axis (
C
) completely. (mechanical close
position).
39. Fasten the plate screws with 2.5 Nm
40. If some grease on plate surface or valve
body clean it with lint and dust free cloth
Allen torque wrench
3mm
Cleanroom wipes
41. Assemble valve unit with control and
actuating. Push axis (
C
) into clamp
coupling
A
B
C