Technical Manual: TM50567
Optica 30/40/50 series
Revision: 4.0
Date of release: 2019-01-23
9/56
www.vareximaging.com
WARNING:
To avoid the risk of electric shock, this equipment must only be connected to
protective earth.
WARNING:
Use of components and accessories other than those specified in this
Technical Manual may lead to unexpected performance of the collimator.
WARNING:
The temperature of the enclosure can become more then 41 °C
CAUTION:
Do not use the collimator if any part is damaged, is making noise or vibrates.
CAUTION:
Do not use the collimator in oxygen-rich environments, and/or in presence of
flammable gases.
CAUTION:
The collimator is not intended for sterilization.
CAUTION:
The collimator is powered through an external power supply at 24 V DC.
CAUTION:
If the enclosure is mechanically damaged DO NOT USE. In case of doubt,
before taking any action, please contact your distributor or Varex Imaging Nederland
B.V..
NOTE:
Always keep this manual and read it before using the collimator.
NOTE:
Please contact your distributor or Varex Imaging Nederland B.V. when the product label
or other markings on the outside of the collimator are missing.
Requirements for SERVICE PERSONNEL
The device may only be installed, repaired or serviced by qualified SERVICE PERSONNEL, i.e.:
•
Authorized by the RESPONSIBLE ORGANIZATION of the device;
•
Authorized by the manufacturer;
•
In possession of the diploma’s / certificates required by national and local legislation.
WARNING:
Service, maintenance and adjustments may only be executed by qualified SERVICE
PERSONNEL.