Technical Manual: TM50567
Optica 30/40/50 series
Revision: 4.0
Date of release: 2019-01-23
37/56
www.vareximaging.com
Guidance and manufacturer’s declaration – electromagnetic immunity – Power and Signal PORTs
The Optica 30/40/50 series is intended for use in the electromagnetic environment specified below. The customer or user of
the Optica 30/40/50 series should assure that it is used in such an environment.
Immunity test
Compliance Test level
Electromagnetic environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV
air
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be at
least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV AC and DC power ports
± 1 kV signal ports
100kHz repetition frequency
Mains power quality should be that of a typical commercial or
hospital environment.
Surges
Line-to-line
IEC 61000-4-5
Not Applicable
The Optica 30/40/50 series is intended to be supplied by a
secondary IEC60601-1 compliant DC power supply.
Surges
Line-to-ground
IEC 61000-4-5
Not Applicable
Conducted
disturbances induced
by RF fields
IEC 61000-4-6
3 V
a)
0,15 MHz – 80 Mhz
6 V
a)
in ISM bands between
0,15 MHz and 80 MHz
b)
80 % AM at 1 kHz
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey
c)
, should be less than the
compliance level in each frequency range.
Interference may occur near equipment marked with the
following symbol:
Voltage dips
IEC 61000-4-11
Not Applicable
The Optica 30/40/50 series is intended to be supplied by a
secondary IEC60601-1 compliant DC power supply.
Voltage interruptions
IEC 61000-4-11
Not Applicable
a)
r.m.s. before modulation is applied
b)
The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553
MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands between 0,15
MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7 MHz to 7,3 MHz, 10,1 MHz to
10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17 MHz, 21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to
29,7 MHz and 50,0 MHz to 54,0 MHz.
c)
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the Optica 30/40/50 is used exceeds the applicable RF
compliance level above, the Optica 30/40/50 series should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating the Optica 30/40/50
series.
NOTE
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people