Technical Manual: TM50567
Optica 30/40/50 series
Revision: 4.0
Date of release: 2018-12-06
46/56
www.vareximaging.com
Functional messages received by the collimator
Description
ID
Byte 1
Byte 2
Byte 3
Byte 4
Byte 5
Byte 6
Byte 7
Set field size in mm
(set in account with SID)
1
Field change
b.0
Cross
b.1
Long
b.2
SID
Cross
MSB
Cross
LSB
Long
MSB
Long
LSB
Spare
SID
0-200
Set shutter position in
byte
(SID is ignored)
2
Field change
b.0
Cross
b.1
Long
Cross
Long
Spare
Spare
Spare
Spare
Set maximum field size
in mm
(set in account with SID)
3
Field change
b.0
Cross
b.1
Long
Cross
MSB
Cross
LSB
Long
MSB
Long
LSB
Spare
Spare
Set max shutter
position in byte
(SID is ignored)
4
Field change
b.0
Cross
b.1
Long
Cross
Long
Spare
Spare
Spare
Spare
Enable max field
(see ID3 and/or ID4)
5
b.0
Max field
off
b.1
Max field
on
Spare
Spare
Spare
Spare
Spare
Spare
Set light and laser
settings
6
Setting
change
b.0
Intensity
b.1
Light Time
b.2
Laser
Time
Light Intensity
Level1 = 20
Level2 = 40
Level3 = 60
Level4 = 80
Level5 = 100
(other values
are ignored)
Light Time
0-60
Laser Time
0-60
Spare
Spare
Spare
Set error status
7
Error Flag
MSB
Error Flag
LSB
Spare
Spare
Spare
Spare
Spare
Set incline value
(This value will overrule the
collimator measured angle
and will be shown on the
collimator touch screen)
8
Incline
MSB
Incline
LSB
Spare
Spare
Spare
Spare
Spare