Guidance and manufacturer’s declaration – electromagnetic immunity –
Compliance Test Levels for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment
Test
frequency
Band
Service
a)
Modulation
Maximum
power
Distance
I
MMUNITY
TEST LEVEL
(MHz)
(MHz)
(W)
(m)
(V/m)
2 450
2 400 – 2 570
Bluetooth;
WLAN
802.11 b/g/n;
RFID 2450;
LTE Band 7
Pulse Modulation
b)
2
0,3
28
217 Hz
5 240
5 100 – 5 800
WLAN 802.11
a/n
Pulse Modulation
b)
0,2
0,3
9
5 500
217 Hz
5 785
a)
For some services, only the uplink frequencies are included.
b)