Technical Manual: TM50567
Optica 30/40/50 series
Revision: 4.0
Date of release: 2018-12-06
48/56
www.vareximaging.com
Description
ID
Byte 1
Byte 2
Byte 3
Byte 4
Byte 5
Byte 6
Byte 7
Configure CAN
messages
All settings will be
stored in the nonvolatile
memory except for the
“Once” message
20
1
ID to change
b.0
ID100
b.1
ID101
b.2
ID102
b.3
ID103
b.4
ID104
b.5
ID105
ID100
0 = Off
1 = Once
Or every
2 = 20 mS
3 = 30 mS
etc. until
255 = 2.55 S
ID101
0 = Off
1 = Once
Or every
2 = 20 mS
3 = 30 mS
etc. until
255 = 2.55 S
ID102
0 = Off
1 = Once
Or every
2 = 20 mS
3 = 30 mS
etc. until
255 = 2.55 S
ID103
0 = Off
1 = Once
Or every
2 = 20 mS
3 = 30 mS
etc. until
255 = 2.55 S
ID104
0 = Off
1 = Once
Or every
2 = 20 mS
3 = 30 mS
etc. until
255 = 2.55 S
ID105
0 = Off
1 = Once
Or every
2 = 20 mS
3 = 30 mS
etc. until
255 = 2.55 S
20
2
ID to change
b.0
ID106
ID106
0 = Off
1 = Once
Or every
2 = 20 mS
3 = 30 mS
etc. until
255 = 2.55 S
Spare
Spare
Spare
Spare
Spare
20
3
See 201, this message wil not influence the nonvolatile memory
20
4
See 202, this message wil not influence the nonvolatile memory
Request Status
message
The status message
will be sent only once
Service Protocol Message
20
5
Message ID of
a status
service
protocol
message
Spare
Spare
Spare
Spare
Spare
Spare
Start calibration cross
shutters
Service Protocol Message
22
0
b.0
Zero
b.1
Max
Spare
Spare
Spare
Spare
Spare
Spare
Start calibration long
shutters
Service Protocol Message
22
1
b.0
Zero
b.1
Max
Spare
Spare
Spare
Spare
Spare
Spare
Start calibration incline
Service Protocol Message
22
2
b.0
0°
b.1
90°
Spare
Spare
Spare
Spare
Spare
Spare