background image

Page 16 

 

                                        3001002 Rev. P    

  WARNING: 

Use high–flow cartridge for flows 5-40 lpm and low-flow cartridge for flows 1-8 lpm.

7-7.  Plug in power cord, and check that all the display indicators light. The Precision Flow

™ 

 

performs a self-test:

 

•  all icons and numeric displays light up for a few seconds

 

•  internal sensors and control systems are checked

 

•  if no faults are detected the unit enters STANDBY mode

 

•  “Water Out” icon indicates there is no water in the disposable water path

 

•  status LED is amber

7-8.  The Precision Flow unit has three controls.
 

Run/Standby Button

 – Powers the unit on and places it in standby.

 

Setting Control Knob

 - Allows you to adjust the parameters.

 

Alarm Mute Button

 – Will intermittently silence alarms and  

 

also dims the display panel. 

 

The Precision Flow has three modes. Those are 

Sleep

,  

 

Standby

, and 

Run

. In Sleep mode, the unit will have a  

 

blank screen and an amber light showing. 

The unit    

 

cannot be started from sleep.

 

To put the unit in 

Standby

, simply rotate the blue  

 

 

Control Setting Knob to illuminate the display. You  

 

 

will see the three parameters of Flow, percent  

 

 

oxygen, and Temperature. There will also be a  

 

 

corresponding cartridge indicator on the lower right  

 

 

hand side which will identify what type of disposable    

 

patient circuit is in place. (Blue/High or Red/Low)

 

To enter 

Run Mode

, with the screen illuminated, 

simply  

 

press and release the Run/Standby Button. 

 

The machine will give a series of 10 beeps, and  

 

begin to power up. At this point the small light  

 

above the Run/Standby Button will change from  

 

Amber to flashing Green. During this start up,  

 

you will also see two amber alarm indicators illuminated.  

 

This is normal and is part of the Precision Flow start up self test. 

7-9.  Push or rotate the control setting knob in either direction to light up the display in  

 

STANDBY mode.

7-10. Press the Mute button to change between bright and dim display  

 

(this function is only available if no alarms are active).

7-11. To connect the sterile water, remove spike cap and wipe 

 

spike with 70-90% isopropyl alcohol. Firmly insert spike  into spike port of the sterile water,  

 

avoiding direct hand contact. Unclamp the water inlet tube so that water (>200 ml) flows  

 

into the disposable water path and the “Water Out” alarm cancels.

7-12. Press Run/Standby button to start gas flow, pump and heater.

  

 

Press twice if the display is initially blank.

  

Section 7 

 Setting up 

Run/Standby  
button

Summary of Contents for precision flow

Page 1: ...Precision Flow Operating Instruction Manual...

Page 2: ...ction 2 Overview 6 Section 3 Principles of operation 7 Section 4 Controls displays connections 8 Section 5 Modes of operation 11 Section 6 Initial assembly 12 Section 7 Setting up 13 Section 8 Adjustm...

Page 3: ...n Emerging and Newly Identified Health Risks SCENIHR has evaluated the exposure to DEHP for the general population and patients during medical procedures In some cases the exposure is significant and...

Page 4: ...ion or maintenance performed by anyone not trained by Vapotherm staff or official training documentation When handling any part of the Precision Flow always follow hospital infection control guideline...

Page 5: ...ounding connection do not operate the device Medical electrical equipment needs special precautions regarding electromagnetic radiation Portable and mobile RF communications equipment can affect medic...

Page 6: ...e adjusted via a single setting control knob on the front panel All values and alarms displayed in a single large color coded panel Flow range 1 40 lpm Oxygen percentage is fully adjustable from 21 to...

Page 7: ...charged The battery is not intended for patient transport gas flow will cease Disposable patient circuit The disposable patient circuit DPC is comprised of the disposable water path DWP vapor transfer...

Page 8: ...ge type 4 Vapor transfer cartridge fault 5 Gas supply fault 6 Status LED 7 Run Standby button see note 8 Setting control knob 9 Alarm mute button 10 Alarm muted LED 11 General fault 12 Water out 13 Bl...

Page 9: ...LED indicates one or more alarms are muted 4 Setting control knob Press to select which variable to adjust Rotate to adjust to new value Press again to set value 5 Hinged door Opens to install or remo...

Page 10: ...nt 3 Access panel for oxygen sensor see note 4 Pole clamp 5 Power cord connection and fuse holder 6 DISS or NIST oxygen connection 7 DISS or NIST air connection 8 Gas inlet filters and traps Note Writ...

Page 11: ...scrub the ports Do not apply organic solvents or bleach Section 5 Modes of operation Mode Action Indicator light color Sleep Display in sleep mode no gas flow Amber Standby Input parameters can be ad...

Page 12: ...installed before first time use The filter and trap assemblies have a quick disconnect fitting which connects to the main unit and a DISS gas fitting for either an oxygen or an air hose Note The quic...

Page 13: ...t not more than 40 102cm above the floor Fixed rail supports may also be used Use with Vapotherm approved IV poles 7 4 Connect oxygen and air supply hoses to correct inlets then connect them to the wa...

Page 14: ...5 2 Fit the delivery tube to the disposable water path as shown Press firmly into place Insert fully Both latches must click shut Assembled disposable patient circuit ready for insertion Water inlet...

Page 15: ...ward as shown 7 6 3 Slide disposable patient circuit downward into the docking station until it stops 7 6 4 Press down firmly to ensure correct seating NOTE If the hinged door does not close easily ch...

Page 16: ...emperature There will also be a corresponding cartridge indicator on the lower right hand side which will identify what type of disposable patient circuit is in place Blue High or Red Low To enter Run...

Page 17: ...tube to stop the flow of water into disposable patient circuit whenever the unit is in standby mode To adjust settings See section 8 Adjustments For alarms and troubleshooting See section 12 Alarms S...

Page 18: ...down unit returns to default settings Section 9 Connecting to patient 9 1 Wait for desired set temperature to be reached before placing the cannula on the end of the patient delivery tube The flashin...

Page 19: ...s at flow rates less than 5 lpm low flow cartridge or less than 10 lpm high flow cartridge To minimize condensation it is recommended not to set the temperature higher than 34 C if using flow rates le...

Page 20: ...es 11 8 Slide the disposable patient circuit into the docking station and close the door 11 9 Hang new sterile water on the IV pole hook 11 10 Wipe the spike on the water inlet tube with 70 90 isoprop...

Page 21: ...E PRIORITIES MEDIUM PRIORITY alarms require immediate attention and are indicated by rapid intermittent tones fast triple beeps LOW PRIORITY alarms require attention as soon as reasonably possible and...

Page 22: ...cannula type re install DPC Water out flashing Medium Priority No water in disposable water path Gas flow continues without heat ing or water circulation Sterile water empty or obstructed inlet tube D...

Page 23: ...path cartridge delivery tube None Cartridge and or DPC not detected STANDBY mode missing cartridge Remove disposable patient circuit Check cartridge installation Cartridge type None Indicates type of...

Page 24: ...o 3003011 should be replaced annually It is accessible by removing a panel at the back of the unit and can be changed in a few minutes by the user or biomedical engineer Use only Vapotherm approved pa...

Page 25: ...orite or 6 maximum Hydrogen Peroxide cleaning solution In addition the following detergent wipes may be used to remove any soil from the unit CaviwipesTM or Sani ClothTM AF3 Germicidal Unplug the Prec...

Page 26: ...state depends on flow rate and temperature Back up power Qty 4 4 8V nickel metal hydride AA batteries not user replaceable Gas supply Medical air and oxygen at inlet pressures between 4 and 85 psi 28...

Page 27: ...ISO 11195 ISTA 2A ENVIRONMENTAL Operating Ambient temperature 18 30 C Ambient relative humidity 0 90 RH non condensing Ambient Pressure Standard atmospheric Not to be used in hyperbaric conditions Sto...

Page 28: ...uter Dia mm Max Flow Premature MN1100A 1 5 8 Neonatal MN1100B 1 5 8 Infant MI1300 1 9 8 Intermediate Infant MI1300B 1 9 8 SOLO cannula SOLO1300 1 9 8 Pediatric Small MPS1500 1 9 20 Pediatric Adult Sma...

Page 29: ...on Flow enters RUN mode from STANDBY when the RUN STANDBY button is pressed Normal operation starts The pump heater and gas flow proportioning systems start Sensors and alarms are active and flow temp...

Page 30: ...ation Power Frequency Magnetics EN 61000 4 8 1993 A1 2001 3A m 50 60Hz PASSED Voltage Dips and Dropouts EN 61000 4 11 2004 Per Standard The Precision Flow uses RF energy only for its internal function...

Page 31: ...rized service center Labor for repairs will not be covered The warranty set forth in this shall become null and void if the Product is opened otherwise tampered with or if repairs are attempted by any...

Page 32: ...Page 32 3001002 Rev P 3001002 Rev P Vapotherm Inc 22 Industrial Drive Exeter NH 03833 USA Phone 603 658 0011 Fax 603 658 0181...

Reviews: