3001002 Rev. P
Page 13
Section 7
Setting up
7-1. Connect power cord if it is not already in place.
7-2. Hang the sterile water from IV pole hook.
7-3. Attach the unit to IV pole below lowest point of
the sterile water.
NOTE
: The Precision Flow
®
oxygen and air supply inlet
fittings are gas-specific to ensure correct connection.
WARNING:
Unit weighs 10.6 lb.(4.81kg) To prevent
possible injury or damage from falling, it must be
securely fixed to a 5-wheel IV pole, with the base of
the unit not more than 40” (102cm) above the floor.
Fixed rail supports may also be used.
Use with Vapotherm approved IV poles.
7-4. Connect oxygen and air supply hoses to correct inlets,
then connect them to the wall outlets.
7-5. Open the bags containing the disposable water path,
cartridge and delivery tube, and assemble them as
follows:
7-5-1. Install a high or low-flow vapor transfer cartridge
in disposable water path as shown. The cartridge
may be inserted either way up. Align the cartridge
ports with the disposable water path openings
and press firmly into place.
Oxygen
Air
Power
cord
Pole
clamp
Disposable
water path
Cartridge
Ports
Cartridge Installed
Disposable
Water path
openings