background image

Page 6 

 

                                        3001002 Rev. P    

Section 2 

  Overview

The Precision Flow® is a system for high flow humidified respiratory therapy by nasal 
cannula. It incorporates the Vapotherm core humidification technology with an electronic 
blender and flow controller. The water and gas pathways are both incorporated into a 
removable, disposable patient circuit.

Features

 

•  The patient circuit is detachable and disposable: no

 

  disinfection necessary   

 

 

•  Minimal downtime between patients: less than  

 

  five minutes to change disposables

 

•  Built-in oxygen/air blender

 

•  Built-in electronic flowmeters and controllers

 

•  Self-testing and self-calibrating

 

•  Internal battery backup maintains flow and    

 

  oxygen percentage for at least 15 minutes if AC  

 

  power is cut off. Battery recharges in 2 hrs.

 

•  All internal sensors self-calibrating and self 

   monitoring

 

•  Single button starts and stops the device

 

•  Temperature, flow and oxygen percentage are  

 

  adjusted via a single setting control knob on  

 

  the front panel

 

•  All values and alarms displayed in a single large  

 

  color-coded panel

 

•  Flow range 1-40 lpm

 

•  Oxygen percentage is fully adjustable from 21 to 100% when two 40 psi (276 kPa) gas  

 

  sources are used

 

•  Inlet gas pressure range is 4-85 psi (28-586 kPa)

 

•  Single gas operation- the Precision Flow

®

 detects inlet gas pressure and blends flow  

 

  based on demand required and available supply. Supply pressure determines FiO

2

 and  

 

  delivered flow; if demand exceeds supply an alarm sounds

 

•  At low gas inlet pressures, maximum flow rate and oxygen percentage settings are    

 

  automatically reduced to match the inlet pressure

 

•  Automatically senses cartridge type: maximum flow setting is automatically reduced if  

 

  low-flow cartridge is installed

 

•  Warm-up time less than five minutes

 

•  Sterile water is connected to the disposable water path using a standard    

 

   spike

 

•  Universal power requirements allow use anywhere with only a change of power cord

 

•  Scheduled maintenance: gas trap filters replaced at 6-month intervals, and oxygen    

 

  sensor replaced annually

Precision Flow

®

Summary of Contents for precision flow

Page 1: ...Precision Flow Operating Instruction Manual...

Page 2: ...ction 2 Overview 6 Section 3 Principles of operation 7 Section 4 Controls displays connections 8 Section 5 Modes of operation 11 Section 6 Initial assembly 12 Section 7 Setting up 13 Section 8 Adjustm...

Page 3: ...n Emerging and Newly Identified Health Risks SCENIHR has evaluated the exposure to DEHP for the general population and patients during medical procedures In some cases the exposure is significant and...

Page 4: ...ion or maintenance performed by anyone not trained by Vapotherm staff or official training documentation When handling any part of the Precision Flow always follow hospital infection control guideline...

Page 5: ...ounding connection do not operate the device Medical electrical equipment needs special precautions regarding electromagnetic radiation Portable and mobile RF communications equipment can affect medic...

Page 6: ...e adjusted via a single setting control knob on the front panel All values and alarms displayed in a single large color coded panel Flow range 1 40 lpm Oxygen percentage is fully adjustable from 21 to...

Page 7: ...charged The battery is not intended for patient transport gas flow will cease Disposable patient circuit The disposable patient circuit DPC is comprised of the disposable water path DWP vapor transfer...

Page 8: ...ge type 4 Vapor transfer cartridge fault 5 Gas supply fault 6 Status LED 7 Run Standby button see note 8 Setting control knob 9 Alarm mute button 10 Alarm muted LED 11 General fault 12 Water out 13 Bl...

Page 9: ...LED indicates one or more alarms are muted 4 Setting control knob Press to select which variable to adjust Rotate to adjust to new value Press again to set value 5 Hinged door Opens to install or remo...

Page 10: ...nt 3 Access panel for oxygen sensor see note 4 Pole clamp 5 Power cord connection and fuse holder 6 DISS or NIST oxygen connection 7 DISS or NIST air connection 8 Gas inlet filters and traps Note Writ...

Page 11: ...scrub the ports Do not apply organic solvents or bleach Section 5 Modes of operation Mode Action Indicator light color Sleep Display in sleep mode no gas flow Amber Standby Input parameters can be ad...

Page 12: ...installed before first time use The filter and trap assemblies have a quick disconnect fitting which connects to the main unit and a DISS gas fitting for either an oxygen or an air hose Note The quic...

Page 13: ...t not more than 40 102cm above the floor Fixed rail supports may also be used Use with Vapotherm approved IV poles 7 4 Connect oxygen and air supply hoses to correct inlets then connect them to the wa...

Page 14: ...5 2 Fit the delivery tube to the disposable water path as shown Press firmly into place Insert fully Both latches must click shut Assembled disposable patient circuit ready for insertion Water inlet...

Page 15: ...ward as shown 7 6 3 Slide disposable patient circuit downward into the docking station until it stops 7 6 4 Press down firmly to ensure correct seating NOTE If the hinged door does not close easily ch...

Page 16: ...emperature There will also be a corresponding cartridge indicator on the lower right hand side which will identify what type of disposable patient circuit is in place Blue High or Red Low To enter Run...

Page 17: ...tube to stop the flow of water into disposable patient circuit whenever the unit is in standby mode To adjust settings See section 8 Adjustments For alarms and troubleshooting See section 12 Alarms S...

Page 18: ...down unit returns to default settings Section 9 Connecting to patient 9 1 Wait for desired set temperature to be reached before placing the cannula on the end of the patient delivery tube The flashin...

Page 19: ...s at flow rates less than 5 lpm low flow cartridge or less than 10 lpm high flow cartridge To minimize condensation it is recommended not to set the temperature higher than 34 C if using flow rates le...

Page 20: ...es 11 8 Slide the disposable patient circuit into the docking station and close the door 11 9 Hang new sterile water on the IV pole hook 11 10 Wipe the spike on the water inlet tube with 70 90 isoprop...

Page 21: ...E PRIORITIES MEDIUM PRIORITY alarms require immediate attention and are indicated by rapid intermittent tones fast triple beeps LOW PRIORITY alarms require attention as soon as reasonably possible and...

Page 22: ...cannula type re install DPC Water out flashing Medium Priority No water in disposable water path Gas flow continues without heat ing or water circulation Sterile water empty or obstructed inlet tube D...

Page 23: ...path cartridge delivery tube None Cartridge and or DPC not detected STANDBY mode missing cartridge Remove disposable patient circuit Check cartridge installation Cartridge type None Indicates type of...

Page 24: ...o 3003011 should be replaced annually It is accessible by removing a panel at the back of the unit and can be changed in a few minutes by the user or biomedical engineer Use only Vapotherm approved pa...

Page 25: ...orite or 6 maximum Hydrogen Peroxide cleaning solution In addition the following detergent wipes may be used to remove any soil from the unit CaviwipesTM or Sani ClothTM AF3 Germicidal Unplug the Prec...

Page 26: ...state depends on flow rate and temperature Back up power Qty 4 4 8V nickel metal hydride AA batteries not user replaceable Gas supply Medical air and oxygen at inlet pressures between 4 and 85 psi 28...

Page 27: ...ISO 11195 ISTA 2A ENVIRONMENTAL Operating Ambient temperature 18 30 C Ambient relative humidity 0 90 RH non condensing Ambient Pressure Standard atmospheric Not to be used in hyperbaric conditions Sto...

Page 28: ...uter Dia mm Max Flow Premature MN1100A 1 5 8 Neonatal MN1100B 1 5 8 Infant MI1300 1 9 8 Intermediate Infant MI1300B 1 9 8 SOLO cannula SOLO1300 1 9 8 Pediatric Small MPS1500 1 9 20 Pediatric Adult Sma...

Page 29: ...on Flow enters RUN mode from STANDBY when the RUN STANDBY button is pressed Normal operation starts The pump heater and gas flow proportioning systems start Sensors and alarms are active and flow temp...

Page 30: ...ation Power Frequency Magnetics EN 61000 4 8 1993 A1 2001 3A m 50 60Hz PASSED Voltage Dips and Dropouts EN 61000 4 11 2004 Per Standard The Precision Flow uses RF energy only for its internal function...

Page 31: ...rized service center Labor for repairs will not be covered The warranty set forth in this shall become null and void if the Product is opened otherwise tampered with or if repairs are attempted by any...

Page 32: ...Page 32 3001002 Rev P 3001002 Rev P Vapotherm Inc 22 Industrial Drive Exeter NH 03833 USA Phone 603 658 0011 Fax 603 658 0181...

Reviews: