Chapter 9 _______________________________________________________________ Maintenance
VAISALA________________________________________________________________________ 79
CHAPTER 9
MAINTENANCE
Cleaning
To ensure the accuracy of measurement results, Weather Transmitter
WXT510 should be cleaned when it gets contaminated. Leaves and
other such particles should be removed from the precipitation sensor
and the transmitter should be cleaned carefully with a soft, lint-free
cloth moistened with mild detergent.
CAUTION
Be extremely careful when cleaning the wind sensors. The sensors
should not be rubbed nor twisted.
Replacing the PTU module
1.
Turn the power off.
2.
Loosen the three screws at the bottom of WXT510.
3.
Pull out the top of the transmitter.
4.
Release the small white flap and remove the PTU
module.
5.
Connect a new PTU module (order code
WXT510PTUSP), replace the top and tighten the
three bottom screws.
6.
Turn the power on.
Release this flap
PTU module
Top of the transmitter