User's Guide _______________________________________________________________________
84 ___________________________________________________________________ M210470EN-B
Rain and wind sensor heating control
The supervisor message
aSU (
see page 75) shows you continuously monitored
information about rain and wind sensor heating (heating temperature Th and
heating voltage Vh).
The heating temperature should stay above 0
°
C when the heating is on (except in
extremely cold conditions where the heating power is not sufficient). The heating
voltage Vh should correspond to the heating voltage supplied. If there is a
remarkable deviation, check the wiring. Note that wire gauge should be large
enough to avoid remarkable voltage drop in the cable.
NOTE
In case AC or full-wave rectified AC is used for the heating, the Vh
measurement behaves as follows:
-
While the heating is off: Vh indicates the positive peak of the heating
voltage waveform
-
While the heating is onVh indicates:
-
mean of the heating voltage waveform with full-wave rectified AC
-
half of the mean of the voltage waveform absolute value with AC.
Operating voltage control
The supervisor message
aSU (
see page 75) shows you continuously monitored
supply voltage level (Vs). In case of deviations between the supplied voltage and
monitored voltage, check the wiring and the power supply.