Chapter 10 ___________________________________________________________ Troubleshooting
VAISALA________________________________________________________________________ 83
Self-diagnostics
Error messaging/Text messages
WXT510 sends a text message when certain type of errors occur. This function
works in all communication modes except in the SDI-12 mode. You may disable
error messaging by using the supervisor message
aSU, S=N,
see page 77.
Examples:
0R1!0TX,Unable to measure error<cr><lf>
(request of wind data while all
the wind parameters were disabled from the wind message)
1XU!0TX,Sync/address error<cr><lf>
(wrong device address.Try the
default address=0)
0XP!0TX,Unknown cmd error<cr><lf>
0xUabc!0TX,Use chksum CCb<cr><lf>
(wrong checksum applied to the 0xU
command).
Text message
identifier (in
NMEA 0183 v3.0
protocol only)
Text Message
Interpretation and action
01
Unable to
measure error
The requested parameters are not activated in
the message, see
Chapter 8
Sensor and data
message settings
, and check the parameter
selection fields.
02
Sync/address
error
The device address in the beginning of the
command is invalid. Ask the device address with
the ?! (SDI-12) or ?<cr><lf> (ASCII and NMEA)
command and retype the command with the
correct address.
03
Unknown cmd
error
The command is not supported, use the correct
command format, see Chapter 7,
Getting the data
messages
.
04
Profile reset
Checksum error in configuration settings during
power-up. Factory settings used instead.
05
Factory reset
Checksum error in calibration settings during
power-up. Factory settings used instead.
06
Version reset
New software version in use.
07
Start-up
Software reset. Program starts from the
beginning.
08
Use chksum xxx
Given checksum not correct for the command.
Use the proposed checksum.
09
Measurement
reset
The ongoing measurement of all the sensors
interrupted and started from the beginning.
10
Rain reset
The precipitation sensor counter reset.
See also troubleshooting, page 81. In case of constant error, please contact Vaisala
Service Center, see page 80.