Figure 28 Insight
Calibrate
menu selection
Restoring factory adjustment
Before adjusting the probe measurement, you can remove any possible earlier adjustments
with the
Restore factory adjustment
selection, if needed.
Figure 29 Insight
Restore factory adjustment
selection
7.2.2 Environmental compensation
When making adjustments, you can enter information about the environment in which you
perform the adjustment. This allows compensating for environmental factors that have an
effect on the measurement (for example, temperature, pressure, or background gases).
The selection of environmental compensations available in the Indigo transmitter
configuration menus depends on the features and configuration of the connected probe. If the
probe you have connected to the transmitter supports environmental compensations, the
configuration options are shown in the
Configure device
menu accessible from the Insight PC
software main view. For information and instructions on the environmental compensation
features of the probe, see the user documentation of the probe you are using.