The service port connection is only intended for temporary use
during configuration, and must not be used for permanent installations. The IP
rating and EMC properties of the transmitter can be affected when the service
port is open.
CAUTION!
Upon connecting, the transmitter receives date and time settings from Indigo80.
1
2
2
3
4
5
OPEN >>
Figure 23 Connecting Indigo300 transmitter to Indigo80 handheld indicator
1
M8 connector of the service cable
2
M12-5F ports on the bottom of Indigo80 for connecting compatible Vaisala devices.
Ports are labeled and .
3
M12-5M A-coded cable connector
4
M12-5F A-coded cable connector
5
Probe (HMP7) connected to Indigo80, for example, for use as a calibration reference
1. Open the protective cover of the transmitter service port using the 4-mm Allen key.
2. Connect the M8 end of the service cable to the service port.
Chapter 5 – Using transmitter with Indigo80 handheld indicator
37
Summary of Contents for Indigo300
Page 1: ...M212849EN A User Guide Transmitter for Vaisala Indigo compatible probes Indigo300...
Page 70: ......
Page 71: ......
Page 72: ...www vaisala com...