Current and voltage test output points
Fixed test output level
Low point test output for voltage output
1 V
High point test output for voltage output
10 V
Always switch the Insight analog output adjustment mode off after
testing or adjusting to return the output to normal operating mode. Normal
measurement data output is not available in the adjustment mode.
CAUTION!
7.2.4.1 Analog output adjustment example
Always switch off the transmitter power supply when connecting or
disconnecting screw terminal wiring. Do not energize the transmitter power
supply when wiring is not connected.
CAUTION!
Always switch the Insight analog output adjustment mode off after
testing or adjusting to return the output to normal operating mode. Normal
measurement data output is not available in the adjustment mode.
CAUTION!
To test and adjust the voltage output level of the transmitter's analog output 1 with Insight:
1. Switch off the transmitter power supply and open the transmitter cover.
2. Connect the multimeter wiring to analog output 1 screw terminals and close the
transmitter cover.
3. Power on the transmitter and connect the transmitter to the Insight PC software.
4. To open the calibration menu, select
to access Insight main menu. In the main menu,
select
Calibrate
.
5. Select
Analog output 1
and enable the adjustment mode with the
Adjustment mode on/
off
selection.
6. Start the low point output test with the
Output low point
button. The analog channel
starts to send the low point test output (1 V in this example).
7. After the low point test output has stabilized, check the multimeter reading. Enter the
reading of the multimeter in the
Measured value, low point
field.
8. Start the high point output test with the
Output high point
button. The analog channel
starts to send the high point test output (10 V in this example).
9. After the high point test output has stabilized, check the multimeter reading. Enter the
reading of the multimeter in the
Measured value, high point
field.
10. Select
Activate adjustment
to save the corrections to the output level.
The corrections are shown in the
Low point offset
and
High point offset
fields.
Chapter 7 – Calibration and adjustment