6. Configuring analog outputs
6.1 Configuring analog outputs with Insight PC
software
6.1.1 Analog output configuration overview
The transmitter has 3 analog outputs that can be configured with the Insight PC software.
Each output has the same set of configuration options in the analog output 1…3 menus under
the
Configure device
menu (
Analog output 1
shown in example).
The available selections depend on the probe that is connected to the transmitter
(for example, CO
2
output parameter selection is only available for CO
2
probes).
Figure 24 Analog output configuration options example in Insight
The
Analog output type
selection in the
General
menu defines whether the outputs are used
for voltage or current output. When you change the output type, Insight reloads the
transmitter.
Indigo300 User Guide
M212849EN-A
40