1. Connect the transmitter to the indicator.
2. Open the main menu by pressing .
3. Select
Devices
. If you have more than one device connected to the indicator, make a
further selection between the devices.
In this case, select the probe connected to the transmitter. For example, if the transmitter
is connected to port 2 of Indigo80, the connected probe is identified with icon
2A
on the
Indigo80 display.
4. If the probe connected to the transmitter supports the sensor purge feature, wait for the
start-up purge to finish or start the purge manually by selecting
Devices > <your device>
> Sensor purge
.
5. Select
Calibrate / Adjust
.
6. Select the parameter, for example,
Relative humidity
.
7. Select
Manual adjustment
.
8. Place the probe to the first reference point and select
Reference value, point 1
to open
the stabilizing graph view. Wait for the measurement to stabilize fully.
a. When the reading has stabilized, press .
b. Use the arrow buttons to set the value of the reference, and then select
OK
or
Set
.
If you already have the reference and measured values available for your
calibration point(s), you can just skip the stabilizing graph view with the
button to get to the view where you can set the reference value. Then replace
the automatically filled
Measured value, point 1
and
Measured temperature,
point 1
with your actual values.
9. If you want to adjust the measurement in 2 points, place the probe to the second
reference point. Then scroll down the
<parameter> adjustment
view and select
Reference value, point 2
to open the stabilizing graph view. Wait for the measurement to
stabilize fully.
a. When the reading has stabilized, press .
b. Use the arrow buttons to set the value of the second reference, and then select
OK
or
Set
.
10. In
<parameter> adjustment
view, check the difference between each reference and
measured value. Very large differences may be due to insufficient stabilization time or
unsuitable calibration setup.
• To store the adjustments to the probe, scroll down and select
Activate adjustment >
OK
. Confirm by selecting
OK
.
• To exit without taking the adjustment in use, press to end calibration.
11. Update the calibration information in
Calibrate / Adjust > Calibration information
.
Navigate to the
Calibration information
menu by pressing .
12. Press to end calibration.
Chapter 7 – Calibration and adjustment
53