Chapter 6 ______________________________________________________________ Maintenance
VAISALA ______________________________________________________________________ 155
Table 31
Error Messages
Error Code Error Message
Action
E0 Humidity
sensor
measurement malfunction.
E1 Humidity
sensor
short
circuit
Check the integrity of the humidity probe and the
probe cable. Clean the probe from dirt, water,
ice or other contaminants.
E2
Humidity sensor open
circuit
E3 Temperature
sensor
open
circuit.
Check the integrity of the humidity probe and the
probe cable.
E4 Temperature
sensor
short
circuit.
E5 Temperature
measurement malfunction
E6 Temperature
sensor
current leak.
Check the integrity of the humidity probe and the
probe cable. Clean the probe from dirt water, ice
or other contaminants.
E7
Internal ADC read error
Internal transmitter failure. Remove the
transmitter and return the faulty unit to Vaisala
Service.
E9
Checksum error in the
internal configuration
memory
E10
Internal EEPROM read
error
E11 Internal
EEPROM
write
error
Internal transmitter failure. Remove the
transmitter and return the faulty unit to Vaisala
Service.
E12 … E13
Add-on module 1 (or 2)
connection failure
Turn off the power and check the module
connection. Turn on the power.
E14 Device
internal
temperature out of range
Ensure that the operating temperature is within
the valid range.
E15
Internal RAM memory or
software checksum error
E18 Internal
ADC
reference
voltage out of range
E19
Internal analog output
reference voltage out of
range
Internal transmitter failure. Remove the
transmitter and return the faulty unit to Vaisala
Service.
E20 … E22
Configuration switches for
analog output 1/2/3 set
incorrectly
Check and re-set the switches, see page 58.
E24 … E25
Internal error in add-on
module 1 (or 2)
Disconnect the power and check the module
connection.
E26 Communication
module
installed in incorrect add-
on module slot
Disconnect the power and change the
communication module to another module slot.
E28 … E29
Unknown/incompatible
module installed in add-on
module slot 1 (or 2)
Ensure that the module is compatible with the
DMT340.
E30
Internal analog voltage out
of range
Internal transmitter failure. Remove the
transmitter and return the faulty unit to Vaisala
Service.
E31
Internal system voltage
out of range
Check that the supply voltage is correct and the
power supply can provide enough power for the
instrument.