Chapter 7 ___________________________________________________Calibration and Adjustment
VAISALA ______________________________________________________________________ 161
4.
Press GRAPH to follow the RH and T stabilization on graphical
display:
0706-020
Figure 83
Following the RH Trend on Graphical Display
5.
Follow the instructions on the display. Use the arrow buttons to
enter the actual humidity of the reference used.
0706-021
Figure 84
Ending Adjustment of Point 1
6.
Continue adjustment of point 2 as instructed by the display:
0706-022
Figure 85
Proceeding to Adjustment of Point 2
7.
Follow the purge progress and wait for stabilization of Point 2 as in
Figure 82 on page 160. Press
READY
when stabilized.
8.
Use the arrow buttons to enter the actual humidity of the reference
used. Follow the instructions on the display to complete the RH
adjustment.
0706-023
Figure 86
Completing Adjustment of Point 2