USER'S GUIDE____________________________________________________________________
126 _________________________________________________________________ M210704EN-F
8.
Select the upper limit
by pressing the
▲▼
arrow buttons. Use the
arrow buttons to set the upper limit value. Press
SET
to confirm
your selection. Press
OK
to confirm your setting.
9. Press
EXIT
to return to the basic display.
AMODE/ASEL
Use the serial line to select and scale the analog output quantities. Check
the analog output modes with the
AMODE
command.
AMODE
<cr>
Example:
>
amode
Ch1 output : 0...1V
Ch2 output : 0...1V
>
Select and scale the quantities for the analog outputs with the command
ASEL
. Note that the optional quantities can be selected only if they have
been selected when ordering the device.
ASEL
[
xxx yyy zzz
]<cr>
where
xxx
=
Quantity of channel 1
yyy
=
Quantity of channel 2
zzz
=
Quantity of the optional analog output channel 3
Always enter all the quantities for all outputs. For quantities and their
abbreviations see Table 2 on page 17.
Use the
ASEL
command as shown in the example below when using a
device with two analog outputs.
Example:
>
asel td t
Ch1 Td low : -20.00 'C ?
Ch1 Td high : 100.00 'C ?
Ch2 T low : -40.00 'C ?
Ch2 T high : 180.00 'C ?
>